Processing math: 100%

2507.07119

Total: 1

#1 First measurement of GaAs as a scintillating calorimeter: achievements and prospects [PDF] [Copy] [Kimi] [REL]

Authors: A. Melchiorre, D. L. Helis, A. Puiu, G. Benato, P. Carniti, I. Colantoni, A. Continenza, N. Di Marco, A. Ferella, C. Ferrari, F. Giannessi, C. Gotti, E. Monticone, S. Nagorny, E. Olivieri, L. Pagnanini, G. Pessina, C. Petrucci, S. Pirro, A. Prajapati, G. Profeta, M. Rajteri, P. Settembri, A. Shaikina, C. Tresca, D. Trotta

In this paper we present the first measurement of a Gallium Arsenide (GaAs) crystal as a scintillating calorimeter with dual heat and light readout within the DAREDEVIL project. The experimental setup features a 4.3 g GaAs (GaAs-1) crystal, operated at approximately 10 mK coupled with a Neutron Transmutation Doped (NTD) thermal sensor for phonon detection and an auxiliary calorimeter for the detection of scintillation light. For the GaAs-1 crystal, a baseline resolution of 121± 2 eV has been achieved. While, with a 3.5 g GaAs (GaAs-2) crystal an even better baseline resolution of 44.5 ± 0.8 eV was achieved. Alpha and X-ray calibration sources were used to study the scintillation light response to different types of interacting radiation. The GaAs crystal exhibits a strong particle discrimination capability based on the emitted scintillation light, featuring a light yield (LY) of 0.9 ± 0.2 keV/MeV for α induced events and 0.07 ± 0.01 keV/MeV for β/γ events, both measured at 1~MeV. The unusual luminescence behavior, i.e. more light being produced under irradiation by α particles warrants further investigation, particularly due to its potential to enhance sensitivity to low-energy nuclear recoils from light dark matter scattering.

Subjects: Instrumentation and Detectors , High Energy Physics - Experiment

Publish: 2025-07-07 16:28:41 UTC