In high-resolution mass spectrometry, an electrospray ionization source is often paired with an ion-funnel to enhance ion transmission. Although it is established that ions experience collision-induced dissociation as they pass through this device, the impact of gas-flow dynamics on ion fragmentation remains unexplored. The present work demonstrates that the gas-flow dynamics from the capillary interface of an electrospray ionization source into an ion-funnel significantly reduces ion fragmentation. This reduction stems from the substantial decrease in the rate of increase in the internal energy of the ions resulting from the collisions with a supersonically expanding gas. The results of this study have significant consequences for systems that employ electrospray mass spectrometry and ion-mobility spectrometry.