Total: 1
4D flow MRI allows for the estimation of three-dimensional relative pressure fields, providing rich pressure information, unlike catheterization and Doppler echocardiography, which provide one-dimensional pressure drops only. The accuracy of one-dimensional pressure drops derived from 4D flow has been explored in previous literature, but additional work must be done to evaluate the accuracy of three-dimensional relative pressure fields. This work presents an analysis of three state-of-the-art relative pressure estimators: virtual Work-Energy Relative Pressure (vWERP), the Pressure Poisson Estimator (PPE), and the Stokes Estimator (STE). Spatiotemporal behavior and sensitivity to noise were determined in silico. Estimators were validated with a type B aortic dissection (TBAD) flow phantom with varying tear geometry and an array of twelve catheter pressure measurements. Finally, the performance of each estimator was evaluated across eight patient cases. In silico pressure field errors were lower in STE compared to PPE, although PPE pressures were less affected by noise. High velocity gradients and low spatial resolution contributed most significantly to local variations in 3D error fields. Low temporal resolution leads to highly transient peak pressure events being averaged, systematically underestimating peak pressures. In the flow phantom analysis, vWERP was the most accurate method, followed by STE and PPE. Each pressure estimator strongly correlated with ground truth pressure values despite the tendency to underestimate peak pressures. Patient case results demonstrated that the pressure estimators could be feasibly integrated into a clinical workflow.