Position-sensitive SiPMs are useful in all light detection applications requiring a small number of readout channels while preserving the information about the incoming light's interaction position. Focusing on a 2x2 array of LG-SiPMs covering an area of 15.5×15.5 mm with just 6 readout channels, we proposed a quantitative method to evaluate image reconstruction performance. The method is based on a statistical approach to assess the device's precision (spatial resolution) and accuracy (linearity) in reconstructing the light spot center of gravity. This evaluation is achieved through a Rice probability distribution function fitting. We obtained an average sensor spatial resolution's best value of 81±3 μm (standard deviation), which is achieved by reconstructing each position with the amplitude of the channels' output signals. The corresponding accuracy is 231±4 μm.

Subject: Instrumentation and Detectors

Publish: 2025-03-07 14:19:50 UTC