Total: 1
Purpose: Inter-scanner variability in cerebral blood flow (CBF) quantification limits reproducibility in multi-center arterial spin labeling (ASL) studies. We developed and validated a vendor-agnostic pseudo-continuous ASL (PCASL) sequence using the open-source Pulseq framework, and assessed its extensibility to advanced perfusion imaging. Methods: A vendor-independent PCASL sequence was implemented in Pulseq and deployed on 3T scanners from Siemens, GE, and United Imaging. Seven healthy volunteers underwent single-post-labeling-delay (PLD) PCASL on all platforms, including test-retest scans. An in-house adapted Siemens vendor-based PCASL sequence served as reference. Reproducibility was assessed using within-subject coefficient of variation (CVwithin), root mean square error (RMSE), regional CBF correlation, and voxel-wise variation ratio maps. Multi-PLD PCASL and planning-free VE-PCASL were implemented as proof-of-concept extensions. Results: Mean gray matter (GM) CBF was similar across implementations. Test-retest GM CVwithin was below 4% on all platforms; Siemens-Pulseq had the lowest value (2.65%), versus 3.64% for the native sequence. GM RMSE between Siemens-Vendor and Pulseq implementations was comparable to test-retest RMSE, and all inter-scanner CVwithin values were below 3%. Voxel-wise variation ratio distributions were centered near 100%, suggesting cross-vendor variability comparable to physiological baseline fluctuations. Multi-PLD and VE-PCASL ran successfully on all platforms, yielding consistent CBF/ATT maps and vascular territory delineations. Conclusion: Pulseq-based PCASL enables vendor-agnostic perfusion imaging with low inter-scanner variability. This open-source approach provides a reproducible, standardized solution for multi-center ASL studies and facilitates deployment of advanced ASL techniques in clinical research.